January 2025
ABSTRACT TITLE: ACCURATE SATURATION MEASUREMENT BY OPTOACOUSTIC MEASUREMENT OF SKIN MELANIN ABSORPTION
Presenting Author: James N Watson, PhD Noninvasix, Inc.;
Co-Authors: David MacLeod, MBBS, Duke University; David Giarracco, Noninvasix, Inc.; Tom Wilmering, Noninvasix, Inc.
Background/Introduction: The effects of skin pigmentation (primarily melanin concentration) on the accuracy of oximetry measurements has been investigated for over 15 years. Historically the disproportionate absorption of red wavelengths was considered to report higher saturation values than those measured by hemoximetry predominantly at lower oxygen (O2) saturation values. Recently, new evidence suggests melanin may introduce a clinically important positive bias at higher saturation values. Inaccurate overestimation of O2 saturation could result in failure to recognize hypoxemia, particularly when remote ‘virtual wards’ are employed, and monitoring occurs in the absence of trained clinical staff. Such risks have prompted a requirement by regulatory authorities for more accurate technologies. Previously, the authors have published evidence of optoacoustic methods for the noninvasive measurement of O2 saturation within targeted vessels. This study investigated the relationship between light absorption by melanin in skin and skin tone using colorimeter measured Individual Typology Angle (ITA). We also investigated whether simultaneous skin tone and O2 saturation monitoring can be used to improve the accuracy of optoacoustic-based saturation measurements. Finally, we provide evidence that epidermal light absorbance changes unpredictably during hypoxia, thereby increasing the need for real-time skin tone compensation.
Methods: During 2022, an IRB-approved, single site, non-randomized, study was executed at Duke University to measure accuracy of an investigational monitor the left innominate vein (LIV) blood saturation, SLIVO2, compared to blood reference values. Over 6 months, 24 healthy, adult volunteers, aged 18 – 45 years old, were enrolled after informed consent. Subjects were selected to obtain a representative range of gender and skin pigmentation. For each subject, the inspired oxygen concentration was lowered in stepwise fashion to produce a controlled oxygen desaturation sequence with SaO2 reported from room air to 70%. At each plateau, serial LIV blood samples were drawn from a central venous catheter placed in the left internal jugular vein and advanced to the LIV and analyzed by a CO-oximeter. The acoustic response to laser illumination of the tissue at 760nm, 800nm and 1064nm were recorded during the breathe-down. Skin tone measurements at the senor site for each subject was measured by colorimeter using the ITA scale and observationally with the Fitzpatrick scale.
Results: Acoustic trace analysis revealed features consistent with the skin and LIV depths identified during prior ultrasound imaging (see figure, top). The observed acoustic peak for different wavelengths at the skin depth was expected to be affected disproportionately by melanin. The median ratio between 760nm and 800nm OA skin peak amplitudes for each subject showed a linear correlation with the measured ITA index (Pearson Corr = 0.75, see figure). However, skin peak ratios were found to not always be stable during the hypoxia breathe-down (see figure). This data indicates a change in fluence at the tissue under skin and suggests a confounding of the saturation measurement at deeper tissue depths. The five subjects (Fitzpatrick Indices: 1, 2, 2, 3, 5) with greatest correlation between their vessel peak amplitude ratios and blood draw saturation were selected for further study. Normalizing vessel peak ratio by skin peak ratio was found to reduce the pooled ARMS error for the population when applying a single calibration curve across all subjects (N = 133; ARMS normalized data = 8.3% ± 16.3 CI95%; ARMS un-normalized data = 19.4% ± 38.1 CI95%). This methodology can be shown to be consistent with Beer-Lambert models applied elsewhere for oximetry measurement.
Conclusion: Optoacoustic signal analysis indicates representative ratios of epidermis absorbance peaks scale linearly with ITA colorimeter measurements. Further investigation demonstrates these skin peak values change over time during controlled hypoxia studies. Compensation consistent with theory reduces ARMS error and supports use of a single calibration curve for SO2 derivation using optoacoustic techniques. This compensation may provide mitigation for the risks associated with skin tone bias during oximetry measurement.